Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Sep 23 | Class II | Mirion Technologies (Capintec), Inc. Device |
Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154; Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground. |